APG-500 Resistance (Pirani) precision transmitter
Purpose of Use:
- The APG500 resistance (Pirani) precision transmitter is designed for vacuum measurement of gases within a vacuum range of 3×10^-2 to 1.0×10^5 Pa.
- It should not be used to measure flammable or explosive gases that can react in certain atmospheres.
- The transmitter must be used with EVP’s control and display instrument or other compatible control and display instruments.
Technical Information:
- Measuring Principle: Based on the physical properties of thermal conductivity of resistance (Pirani).
- Regulatory Interface: DN 16 ISO.
- Measuring Range: Suitable for air, nitrogen, oxygen, and carbon monoxide within the range of 3×10^-2 to 1.0×10^5 Pa.
- Accuracy: Varies based on the pressure range, with a maximum accuracy of ±30% of reading.
- Repeatability: 5% of reading within certain pressure ranges.
Output Signal (Measurement Signal):
- Voltage Range: 0 to +9.5 VDC.
- Output Voltage: +2.5 to +9.0 VDC.
- Voltage Vacuum Ratio Logarithm: 1.0V per level.
- Error Signal: 0 to +0.5VDC (filament broken).
- Output Impedance: 2×5.1 Ω.
- Response Time: 100ms.
- Control Range: 1×10^-1 to 5.0×10^3 Pa.
- Control Output: 1 open collector transistor output (DC30V/100mA).
- Transmitter Identification: 27.0 kΩ.
Power:
- Transmitter Input Voltage: +13.5 to +30 VDC.
- Ripple: ≤5% Vpp.
- Current: <300mA (startup current).
- Power Consumption: ≤2 W.
- Cable Connection: FCC 68 / RJ45 electrical connector, 8-pin electrode.
- Transmitter Cable: 8-core cable, length ≤100 meters.
Safety and Installation:
- The transmitter must only be connected to a power supply, instrument, or control instrument that meets the requirements of grounded ultra-low voltage.
- Power connection must be equipped with a fuse.
- Installation direction: Vertical installation.
Materials and Environment:
- Materials exposed to vacuum include tungsten, nickel, stainless steel, and nitrile rubber.
- Allowable pressure: 0.3 MPa (absolute pressure), limited to inert gas.
- Allowable temperature for vacuum measurement: +5 to +45°C; Vacuum connector: 80°C; Sensing wire: 120°C; Storage: -5 to +60°C.
- Relative humidity: <80% at temperatures up to +31°C, reduced to 50% at +40°C.
- Intended for indoor use only.
Danger:
- Ensure proper grounding and adhere to safety guidelines to prevent hazards associated with electrical connections and operation.
For more information about the APG-500 Resistance (Pirani) precision transmitter, contact us
WhatsApp: +86 13849062209
Our team is ready to provide you with professional service and support.
Keywords
vacuum system,nuoxutech, vacuum pump, VacuumSystem, APG-500 Resistance (Pirani) precision transmitter.